Choosing a lab incubator comes down to two numbers: the temperature range your protocol demands and the chamber volume your workload fills. This page sets out the ZSH series — a bidirectional 0–70 ℃ laboratory incubator with ±0.5 ℃ constant-temperature deviation — so you can size, specify and quote in one pass.

Digital display biochemical incubator for microbial cultivation and plant tissue culture

Start with the number you already know

Route by temperature control range

Most protocols are decided by a single fact: whether the chamber must pull temperature below the room or only hold above it. Pick the condition that matches your method and jump straight to the specification.

The ZSH series controls both directions on one compressor circuit, so a run that starts at 4 ℃ and finishes at 37 ℃ needs no second chamber.

ZSH series biochemical incubator cabinet with large LCD screen and touch buttons

Series overview

One refrigerated family, 70 L to 1500 L

The ZSH series is a single platform rather than a set of separate products: the same controller, the same refrigeration module and the same chamber construction scale from a benchtop footprint up to a floor-standing cabinet.

0~70 ℃Bidirectional cooling and heating control range
±0.5 ℃Constant temperature deviation
±1.5 ℃Temperature fluctuation, no load @25 ℃
0.1 ℃Temperature resolution on the LCD display
  • Microcomputer PID controller with a large LCD screen and touch-button setting for temperature and timing duration.
  • Brand refrigeration compressor on eco-friendly refrigerant, with start delay and shutdown interval protection.
  • Cross-flow fan generating horizontal circulation across the working area.
  • Universal casters with independent support feet for repositioning and levelling.

Specification matrix

Master specification comparison

Every figure below is published on the ZSH product page. Use the deviation and fluctuation rows to compare against your method's tolerance, and the dimension rows to check the unit against your bench or doorway clearance.

Specification Published value What it decides for you
Temperature control range0 ~ 70 ℃Whether sub-ambient and above-ambient holds can share one chamber
Constant temperature deviation±0.5 ℃How close the setpoint sits to the value your method quotes
Temperature fluctuation±1.5 ℃ (no load @25 ℃)Expected swing around the setpoint once the chamber is empty
Temperature resolution0.1 ℃Smallest step you can read and set on the display
Timing range1 ~ 9999 minLongest unattended run you can program in one entry
Circulation modeHorizontal cross flow circulation fanHow air moves across shelves rather than straight down onto samples
Inner materialDetachable stainless steel linerCleaning and disinfection routine between batches
Safety functionsOvertemperature audio & visual alarm, power-off parameter memory, compressor start delay protectionRecovery behaviour after an alarm, a power cut or a restart
Power supplyAC220V on small-volume models; AC380V on extra-large capacity modelsWhether your lab needs a three-phase supply
Model Power supply Input power Inner W×D×H (mm) Overall W×D×H (mm) Nominal volume Standard load carriers
ZSH-70 / ZSH-70FAC220V 50Hz500 W450×320×500580×550×110070 L2 pcs
ZSH-150 / ZSH-150FAC220V 50Hz900 W480×400×780605×625×1375150 L3 pcs
ZSH-250 / ZSH-250FAC220V 50Hz1150 W580×500×850705×725×1555250 L3 pcs

Published dimension data covers ZSH-70 through ZSH-250. The series continues through 300 L, 350 L, 500 L, 800 L, 1000 L and 1500 L — send your target volume and we will confirm the corresponding inner and overall dimensions, load carrier count and supply requirement before you order.

Chamber build

What the door, liner and airflow do for your data

Three construction details change how often your readings drift and how long a batch changeover takes.

  • Double-layer hollow tempered glass door. Samples stay visible without opening the chamber, and a thermal sealing strip limits heat exchange so temperature recovers quickly after the door is opened.
  • Detachable stainless steel liner. The liner comes out for cleaning and disinfection, removing dead corners and cutting the risk of microbial cross-contamination between runs.
  • Horizontal cross-flow circulation. The cross-flow fan drives air across the working area rather than onto one spot, which is what makes a shelf-to-shelf uniformity claim possible.
  • Coated cold-rolled steel shell with angular-arc door corners. Rounded internal corners leave nowhere for residue to collect.
Biochemical incubator control panel and chamber controls
Floor-standing biochemical incubator with castors and independent support feet

Capacity and power

Volume band decides the footprint and the supply

Volume is not only a bench-space question. It also changes the input power, at the top of the range, the electrical supply your facility has to provide.

Volume bandPublished referencePlanning note
70 LZSH-70: 580×550×1100 mm overall, 500 W, AC220V 50HzBenchtop-scale batch, 2 load carriers
150 LZSH-150: 605×625×1375 mm overall, 900 W, AC220V 50Hz3 load carriers, still on a standard wall socket supply
250 LZSH-250: 705×725×1555 mm overall, 1150 W, AC220V 50HzTallest published AC220V unit at 1555 mm — check doorway height
300–1500 LSeries volumes listed up to 1500 LExtra-large capacity models adopt AC380V — confirm the supply before electrical work is scheduled

Universal casters let the cabinet move to its final position; independent support feet then lock and level it so the chamber does not rock during door operation.

Scenario to specification

Match your method to a configuration

Four common laboratory scenarios, each translated into the specification that carries the decision — plus the one line to confirm on the quotation.

ScenarioSpecification that decides itConfiguration to confirm
Five-day BOD series in a warm laboratory Sub-ambient capability plus a powered dissolved-oxygen probe 0~70 ℃ cooling/heating chamber with the waterproof BOD test socket option
Pharmaceutical QC incubation at 37 ℃ ±0.5 ℃ constant-temperature deviation and a programmable run length ZSH with 1~9999 min timing; add the micro printer for a paper record
Seed germination and plant tissue culture Horizontal airflow across the working area and easy disinfection between batches Detachable stainless steel liner with cross-flow circulation; 150 L or 250 L
Unattended overnight or weekend runs Recovery after a power interruption Power-off parameter memory plus compressor start delay protection
Biochemical incubator chamber detail showing the observation door and stainless steel interior

Optional accessories

Which option answers which problem

Options are usually the difference between a chamber that fits the method and one that just holds temperature. Each row pairs an accessory with the problem it removes.

Biochemical incubator with interior shelving and accessories visible inside the chamber
OptionProblem it answers
Continuously variable speed fanSample water evaporation is skewing your results — wind speed can be dialled down
BOD test socketYou need a waterproof socket to power a dissolved-oxygen probe inside the chamber
Independent temperature limiting controllerYour SOP requires a second, independent over-temperature cut-off
RS485 communication interfaceChamber readings have to reach a building system or records platform
Micro printerEach run needs a hard-copy record at the end of the cycle
Colourful touch screenOperators want a richer interface than the standard LCD with touch buttons

Options are specified at order stage and fitted to the same cabinet — list the ones your method requires when you request the quotation.

Send your spec, we confirm the model

Tell us the temperature range, the volume band your batch fills, your uniformity tolerance and any accessories your method requires. We will come back with the matching ZSH configuration, the dimension drawing for your chosen volume, the supply requirement and a quotation.

Related Reading

Three notes to help clarify product requirements before contacting the supplier.

Reading Incubator Supplier Specifications in Product Listings

A laboratory incubator supplier listing is a published set of product data, and every field makes sense only with the condition printed beside it.

Read article

BOD Incubation and Microbial Oxygen Demand in Water Testing

BOD testing ties three things together at once — the organic matter in a water sample, the oxygen microbes consume while breaking it down, and the incubation conditions that keep that consumption…

Read article

How Does a Laboratory Incubator Hold a Steady Temperature?

A laboratory incubator holds a steady temperature by balancing heat added by its heater against heat escaping through walls, door, gasket, and load, with a controller and air path managing that…

Read article

Frequently Asked Questions

what temperature range does the zsh biochemical incubator control?

the zsh series controls 0 to 70 ℃ in both directions through a bidirectional cooling and heating system, so it can hold sub-ambient and above-ambient setpoints in one chamber.

how stable is the temperature inside the zsh chamber?

the published constant temperature deviation is ±0.5 ℃, temperature fluctuation is ±1.5 ℃ with no load at 25 ℃, and the display resolves temperature to 0.1 ℃.

which chamber volumes are available in the zsh series?

zsh volumes include 70 l, 150 l, 250 l, 300 l, 350 l, 500 l, 800 l, 1000 l and 1500 l. published model dimensions cover zsh-70, zsh-150 and zsh-250; confirm larger sizes before ordering.

how does airflow affect temperature uniformity across the shelves?

a cross-flow fan creates horizontal circulation inside the chamber, moving air across the working area. an optional continuously variable speed fan lets you reduce wind speed and sample water evaporation.

what happens after a power cut or an overtemperature event?

the controller keeps parameter memory and power-failure data recovery, sounds an audio and visual overtemperature alarm, and uses compressor start delay protection to reduce frequent start-stop wear.

which optional accessories can be fitted to the zsh incubator?

available options include a continuously variable speed fan, a bod test socket, an independent temperature limiting controller, an rs485 communication interface, a micro printer and a colourful touch screen.

labcarta Technology Co., Ltd.Ilabcarta Technology Co., Ltd.

CONTACT US

If you have any queries, get in touch today! Don't hesitate. We try to take the extra step for our customer satisfaction.
Name *
Email *
Whatsapp/Phone
Message *
Verification Code *
Click to verify
Leave a message
Name *
Email *
Whatsapp/Phone
Message *
Verification Code *
Click to verify
We use Cookie to improve your online experience. By continuing browsing this website, we assume you agree our use of Cookie.